by Richard A. Mentelos, RAM Technologies, LLC, Vendor website, Feb 01 2007
Focus:
Generic computer power supplies such as AT or ATX types are generally unsafe for use in a
medical environment because of high leakage current and improper insulation as defined by
IEC601-1 and UL2601-1. To comply with leakage current requirements requires either the addition
of an isolation transformer or the use of a medical-grade power supply (the author makes the
case for the latter.) Leakage current requirements for UL2601-1 and IEC601-1 are discussed
including the differences in the two standards as well as the changes in leakage current
requirements from the second edition of IEC601-1 to the third edition of this standard. Hi-POT
and EMC requirements are also discussed.
What you’ll learn:
Notes:
RAM Technologies Application Note AN-11
View this Source (requires a PDF Viewer installed on your device)